RGM180 · QUICK START GUIDE
RGM180 Display Series quick start guide
Quick Start Guide RGM180 Display Series
Transcribed from the SATEC document RGM180 Display Series Quick Start Guide, BG0509 REV.A2, April 2012 (PDF, 24 pages, 1492 KB). Drawings and diagrams are in the PDF. If this page and the PDF differ, the PDF is right; tell us through the contact form.
LIMITED WARRANTY
The manufacturer offers the customer a 24-month functional warranty on the instrument for faulty workmanship or parts from date of dispatch from the distributor. In all cases, this warranty is valid for 36 months from the date of production. This warranty is on a return to factory basis.
The manufacturer does not accept liability for any damage caused by instrument malfunction. The manufacturer accepts no responsibility for the suitability of the instrument to the application for which it was purchased.
Failure to install, set up or operate the instrument according to the instructions herein will void the warranty.
Only a duly authorized representative of the manufacturer may open your instrument. The unit should only be opened in a fully anti-static environment. Failure to do so may damage the electronic components and will void the warranty.
The greatest care has been taken to manufacture and calibrate your instrument. However, these instructions do not cover all possible contingencies that may arise during installation, operation or maintenance, and all details and variations of this equipment are not covered by these instructions.
WARNING
All trademarks are property of their respective owners.
April 2012
Copyright 2003-2007 © SATEC Ltd.
General
The RGM180 series provides a user-friendly interface to monitor and configure SATEC devices.
The RGM180 series includes two models:
- Local Graphic Display RGM180-G1: the RGM180 communicates with a single host processor expertmeter™ SATEC device, using an RS-485 interface. It is powered directly from the device.
- Network Graphic Display RGM180-G3: the RGM180 communicates with multiple host processor expertmeter™ SATEC devices (up to 32), using a 10/100Base T Ethernet or RS-485 interface. It is powered by Power over Ethernet (PoE) and/or an external AC/DC adaptor.
Installation: 1. Interface Description
Front and rear views of the RGM180 with numbered callouts 1 to 6. Drawing not reproduced on this page. See page 3 of the PDF.
- LED: Red (Wh)/Green (Ethernet Link).
- IR Communication port (for PM180 only)
- USB Communication port
- RS485 Communication port and PM17x/PM180 display connector
- Ethernet Communication port (RGM180-G3 only)
- Display & Touchscreen.
2. Connection Configurations: 2.1 PM17x / RPM17x
Figure 1: PM17x connection. Drawing not reproduced on this page. See page 4 of the PDF.
Figure 2: RPM17x Connection. Drawing not reproduced on this page. See page 4 of the PDF.
2.2 ezPAC Connection
Figure 3: ezPac Connection. Drawing not reproduced on this page. See page 5 of the PDF.
2.3 EM720 Connection
Figure 4: EM720 Connection. Drawing not reproduced on this page. See page 5 of the PDF.
2.4 PM130 PLUS Connection
Figure 5: PM130 Plus Connection. Drawing not reproduced on this page. See page 6 of the PDF.
2.5 EM133 Connection
Figure 6: EM133 Connection. Drawing not reproduced on this page. See page 6 of the PDF.
2.6 Ethernet Connection
2.6.1. Regular Ethernet connection
Figure 7:Regular Ethernet Connection. Drawing not reproduced on this page. See page 7 of the PDF.
2.6.2. PoE Ethernet Connection
Figure 8: PoE Ethernet Connection. Drawing not reproduced on this page. See page 7 of the PDF.
2.7 Mechanical Installation
Figure 9: Instrument Dimensions. Drawing not reproduced on this page. See page 8 of the PDF.
2.8 RGM180 Panel Mount Installation
Figure 10: Panel Mount Installation. Drawing not reproduced on this page. See page 9 of the PDF.
Operation: 1. Connection
Connecting to a Meter
The RGM runs a self-test prior to connection.
The screen will display “Connecting….;” when connected, Voltage measurement data will be displayed.
Screen image: Voltage Line to Line screen. Drawing not reproduced on this page. See page 10 of the PDF.
The screen is touch activated.
To display the Main Menu, press MENU 3 times.
The Main Menu allows you to:
- Select the data group you wish to be displayed (see Section 6);
- Conduct display and instrument setups (see Sections 2 and 3).
Screen image: Main Menu. Drawing not reproduced on this page. See page 10 of the PDF.
Connecting to the RGM
The device allows configuration through the USB or Ethernet at Modbus address 99.
2. Display Setup
Press on Display Setup from the Main Menu. The Display Setup menu will appear.
Screen image: RGM Setup Menu. Drawing not reproduced on this page. See page 10 of the PDF.
Display Definitions
Press Display to set Brightness, Standby Mode and Language:
After setting these options, press Save and then close the window by pressing on the X in the top right corner.
Screen image: Display Setup window. Drawing not reproduced on this page. See page 11 of the PDF.
Screen Calibration
Press Calibration to calibrate the display touch screen sensor:
Follow the instructions on the screen to calibrate the display. A message will appear indicating if input values are out of range.
Screen image: touch screen calibration. Drawing not reproduced on this page. See page 11 of the PDF.
Press OK to return to Display Setup menu.
Network Setup
Press Network to define the network connection:
Press on the IP Address number to define the address; input the correct IP address (or press Clear to input a new address) and press OK to save or Cancel to restore the previous address.
Define or change the Subnet Mask number and Default Gateway in the same manner.
DHCP – Press the setting (Off or On) to change, then press OK to save.
Press X to close the window and return to the Display setup menu.
Screen image: Network Setup window. Drawing not reproduced on this page. See page 11 of the PDF.
Serial Port Settings
Press Serial to define the serial port settings:
Set the Baud rate, Data format, Port direction and response time in the same manner as in the Network Setup.
Press X to close the window and return to the Display setup menu.
Screen image: Serial Port window. Drawing not reproduced on this page. See page 12 of the PDF.
Factory Settings
Press Factory to view the factory-set definition of the RGM.
Screen image: Factory window. Drawing not reproduced on this page. See page 12 of the PDF.
Press MENU to return to the Main Menu.
Screen image: Main Menu. Drawing not reproduced on this page. See page 12 of the PDF.
3. Instrument Setup
The RGM connects to up to 32 devices. Each device has its own instrument setup definition.
Press Instrument Setup to define Wiring Setup and Demand Setup, to adjust the Time, to Reset measurement parameters and to define Communication parameters.
Screen image: Instrument Setups menu. Drawing not reproduced on this page. See page 13 of the PDF.
Wiring Setup
Press Wiring Setup to define the Wiring Mode, CT primary and Nominal Frequency.
Press on the value of the desired parameter and enter the correct value. Press Receive to complete the definition, then close the window using the X in the upper right corner.
Screen images: Wiring Setup window and Wiring Mode selection. Drawing not reproduced on this page. See page 13 of the PDF.
Demand Setup
Power demand parameters can be defined on this screen, used in the same fashion as the Wiring Setup.
Time Adjustment
Use this screen to set the date and time. Press OK when completed. Press OK on the ‘Operation successfully completed’ message.
Reset Setup
Use this screen to reset parameters or clear logs. You will be asked after each operation if you are sure you want to perform the reset.
Communications
Communications parameters can be set in the same fashion as the Wiring and Demand Setups.
Press MENU to return to the Main Menu.
Screen image: Main Menu. Drawing not reproduced on this page. See page 14 of the PDF.
4. Data Display
From the Main Menu, press RT Measure to display the Select Data Menu.
The Select Data Menu allows you to choose from the four data groups: Real Time Measurements, Harmonics, Energy, and the Logs (Event Log).
To display Real Time Measurements, press on Real Time.
Screen image: Select Data menu. Drawing not reproduced on this page. See page 15 of the PDF.
4.1 Real Time Measurements
Press on the desired data item to display the data. Continue to press NEXT to progress through the data in this group:
- Voltage: RMS, minimum, maximum and maximum demands
- Current: RMS, minimum, maximum and maximum demands
- Power: Total, minimum, maximum, maximum demands, power per phase
- Auxiliary power: Total, minimum, maximum
- Phasor and waveform graphs
Screen image: Real Time Data menu. Drawing not reproduced on this page. See page 15 of the PDF.
This screen can be reached at any point from any of the real time measurement screens by pressing on MENU from those screens.
Press Volts L-L RMS to display the screen at right.
Press on L-N at the bottom of the screen to view Line to Neutral voltage; press on Sec to display Line to Line Secondary voltage.
Press NEXT to proceed to the next data item.
Screen image: Voltage Line to Line screen. Drawing not reproduced on this page. See page 15 of the PDF.
To maneuver between the data items displayed, press PREV or NEXT.
At the display of voltage maximum demand, press NEXT to display Current data.
Current (Amps)
The next four screens display:
- Currents (Amps)
- Minimum currents
- Maximum currents
- Maximum Demand currents
Screen image: Amps screen. Drawing not reproduced on this page. See page 16 of the PDF.
Press NEXT to display Power data.
Power
The next five screens display:
- Total Power: kW, kvar, kVA, PF
- Total Power Minimum
- Total Power Maximum
- Power Demands: Maximum, Present and Predicted
- Power per phase
Screen image: Power Total screen. Drawing not reproduced on this page. See page 16 of the PDF.
Press NEXT to display Auxiliary power data.
Auxiliary Power
The next three screens display:
- Auxiliary power
- Minimum auxiliary power
- Maximum auxiliary power
Screen image: Auxiliary screen. Drawing not reproduced on this page. See page 16 of the PDF.
Press NEXT to display the phasor graph.
Phasor
The phasor is a graphical display of phase angles for voltage and current per phase.
The phases are color coded for easy reading of the graph.
Screen image: phasor graph. Drawing not reproduced on this page. See page 17 of the PDF.
Press NEXT to display the Waveform graph.
Waveform
The waveform is a real time graphical display.
Use the sliding bar on the right to view different waveforms.
The icons at the top are (from left to right):
- Channel selection
- Zoom
- View previous or next section of the graph
- View waveform properties
Screen image: RT Wave Form screen. Drawing not reproduced on this page. See page 17 of the PDF.
Press MENU to close this screen. “Closing” will appear, after which you will be back to the Real Time Measurement Data Menu (see 4.1).
From the Real Time Measurement Data screen, press MENU to display the Select Data menu; press NEXT to display the Harmonic Data Menu (see 4.2).
Individual Harmonics
These screens work differently from the Real Time Measurement Data screens – you will always need to press MENU to return to the screen shown above and then select the next item you want to view.
Each harmonic spectrum can be viewed in graph or tabular form, as shown below. The table expresses the harmonic distortion as a percentage. To view the full table, use the scroll bar on the right.
Press on V1 to display the Voltage Harmonic Spectrum (displayed per phase) graph:
Screen image: V1 Spectrum graph. Drawing not reproduced on this page. See page 18 of the PDF.
Press on the table icon at the bottom of the screen to view the Voltage Harmonic Spectrum table.
Screen image: V1 Spectrum table. Drawing not reproduced on this page. See page 18 of the PDF.
View Current Harmonic Spectrum (per phase) in the same way, by pressing on I1 on the Harmonic Data screen above.
Total Harmonic Distortion
These screens work the same as the Real Time Measurement Data screens – press NEXT to proceed to the next data item, or PREV to return to the previous data screen.
Voltage THD – per phase
- Total
- Minimum
- Maximum
- Maximum Demands
Current THD – per phase
- Total
- Minimum
- Maximum
- Maximum Demands
Screen image: Voltage THD screen. Drawing not reproduced on this page. See page 19 of the PDF.
Press MENU to return to the Harmonic Data group screen. Press NEXT to proceed to the Energy Data Menu (see 4.3).
4.3 Energy Data
This screen presents the Energy data group, consisting of the following data:
- Energy import: kWh, kvarh, kVAh
- Energy export: kWh, kvarh
- Net energy
Screen image: Energy Data menu. Drawing not reproduced on this page. See page 19 of the PDF.
Press on the desired data item to view that data.
Screen image: Energy Import screen. Drawing not reproduced on this page. See page 19 of the PDF.
Press NEXT or PREV to see next or preceding data item. Press MENU to return to Energy Data Menu.
Press MENU to return to Select Data Menu; Press Logs to view Log Selection screen (only Event Log is available).
4.4 Log Selection
Press Event Log to define and view the log.
Screen image: Log Selection screen. Drawing not reproduced on this page. See page 20 of the PDF.
When you start the time, the Log will download to the display.
To order the data, first define the start time. Press on the gray area to insert the time definition. Press OK. Press OK again to commence log downloading.
Screen image: Event Log, Order the Data window. Drawing not reproduced on this page. See page 20 of the PDF.
The log will appear. Use the scroll bars at bottom and right to view the full log. To re-start the log, press READ to again define the start time.
Screen image: Event Log table. Drawing not reproduced on this page. See page 20 of the PDF.
Press MENU to return to Log Selection Menu.
Press MENU again to return to Select Data Menu.
5. Meter Selection
Labels
The label name assigned to a device will appear in the lower left of the screen. The user may change these label names (the actual device being displayed will appear at the top of the screen in parentheses).
Meter Selection
To change the connection to a different device, press on the label name in the lower left of the screen (in this screen, it is PM172).
Screen image: Voltage Line to Line screen. Drawing not reproduced on this page. See page 21 of the PDF.
The Meter Selection screen will appear.
Press on the address of the desired meter. Press OK on the ‘Connection established’ box. Data display will begin.
Screen image: Meter Selection screen. Drawing not reproduced on this page. See page 21 of the PDF.
To change the label name or address of a device, press on Setup. The Connection Manager screen will appear. Press on the label name or address you wish to change.
Screen image: Connection Manager screen. Drawing not reproduced on this page. See page 21 of the PDF.
To perform display setup from this menu, press RGM Setup. See Section 2 for Display Setup instructions.
6. Troubleshooting
- If screen does not respond to touch – perform screen calibration (see Section 2, Display Setup).
- If the display does not connect to the remote meter – make sure that the Modbus and/or IP address specified in the Connection Manager (see Section 5) match the remote meter address.
- If the display does not connect through the Ethernet connection – make sure that the installed IP address is set correctly (see Section 2, Display Setup > Network Setup). You may also need to contact your system administrator to obtain a legal address in a specific network.
Technical Specification
| COM1 | IR Communication port | PM180 device Only |
| Optional | Optical Communication port | |
| Optional | Max. Baud rate | 19.200 kb/s |
| Optional | Protocols | Modbus or DNP3.0 |
| Optional | Isolation | 2500 VAC @ 1 mn |
| COM2 | Serial Communication port | Device COM port |
| basic | RS-485 Max. Baud rate | 115.2 to 480 kb/s |
| basic | Isolation | 4000 VAC @ 1 mn |
| basic | RS-485 Maximum length cable | 1000m |
| basic | Protocols | Modbus RTU |
| basic | Connection | DB-15 |
| ETHERNET | Multiple device Communication port | RGM180-G3 only |
| 10/100Base-T Basic | Built-in network communication port with PoE ability | |
| 10/100Base-T Basic | Wired LAN communication port with auto-negotiation | IEEE 802.3 |
| 10/100Base-T Basic | Ethernet port Baud rate | 10/100 Mb/s |
| 10/100Base-T Basic | Protocols | Modbus/TCP |
| 10/100Base-T Basic | ETH port Isolation | 1.5 KVAC @ 1mn |
| 10/100Base-T Basic | ETH connector | Standard RJ-45 |
| USB | Built-in USB Communication port | DISPLAY Panel |
| USB | Basic - Device (default): USB communication port | Full speed Device |
| USB | Basic - Device (default): USB port Baud rate | 12 Mb/s |
| USB | Basic - Device (default): Protocols | Modbus RTU/ASCII and DNP3.0 |
| USB | USB device port Isolation (Remote display) | 1500 KVAC @ 1mn |
| USB | USB device port Isolation (Local display) | 4000 KVAC @ 1mn |
| USB | USB connector DISPLAY Panel | USB type A, vertical mount, straight |
| size | 5.7″ |
| resolution | 320 x 240 dots |
| Type | TFT – color with Touch Panel |
| Outline dimensions | 131mm (W) x 102mm (H) x 14.5mm (D) |
| Active area | 115.2mm (W) x 86.4mm (H) |
| Operating temperature | -20°C - +70 °C |
| Storage temperature | -30°C - +80 °C |
| For energy and tariff registers logging, EV-PQ-DATA-WV log | Basic 256MB |
Power supply
| 12VDC – Device PS standard | Rated Input | 10.8 – 13.2V DC |
| 12VDC – Device PS standard | Dielectric withstand insulation | 4000 VDC @ 1mn |
| 12VDC – Device PS standard | Power Consumption | 2W |
| 24VDC - Device PS option (PM180 Aux. PS) | Rated Input | 9.5 – 24V DC |
| 24VDC - Device PS option (PM180 Aux. PS) | Dielectric withstand insulation | 3000 VDC @ 1mn |
| 24VDC - Device PS option (PM180 Aux. PS) | Power Consumption | 2W |
| 48VDC – PoE option | Rated Input | 37 – 58V DC |
| 48VDC – PoE option | Dielectric withstand insulation | 1500 VDC @ 1mn |
| 48VDC – PoE option | Power Consumption | 2W |
| All models | Detachable Terminals for wires size | 3 x 2.5 up to 6 mm² |
| All models | Header pitch | 7.5 mm |
| All models | PoE connection | RJ45 |
| Operational temperature | -30 °C to 70 °C |
| LCD Operational temperature | -20 °C to 70 °C |
| Storage temperature | -30 °C to 80 °C |