PM175 ยท CT PRIMARY

What does the PM175 LED display's CT primary setting mean?

SATEC PM175 power quality analyser, front panel with load bar and three rows of red LED digits
PM175 product photograph from the manufacturer's manual cover

On a PM175 with the three-row LED display, the Basic setup menu has an item labelled Ct, "CT primary current". This page explains what the manual says it means and how to view it without changing it.

It applies to the established PM175 with the LED display and standard 1 A or 5 A CT inputs. It does not cover the PM175 TFT touch-screen display, the PM175 PRO, or the HACS, RS5, SC5 and SC50 current-input versions.

What the manual says Ct is

The Basic setup table in the PM175 manual lists Ct as follows.

LabelParameterOptionsDefaultDescription
CtCT primary current1-20,000 A5 AThe primary rating of the phase current transformer

In the same table, the PT ratio setting (Pt) has the description "The phase potential transformer's primary to secondary ratio". The table also lists ConF (wiring connection mode), Pt.F (PT ratio multiplier) and U.SEC (nominal secondary voltage).

The CT secondary is a separate choice. The PM175 is ordered with 5 A or 1 A current inputs. In PAS, the manual's instruction is to "Select a correct CT secondary current (5A or 1A) for your meter" when setting up the site.

The manual gives two cautions with this table:

  • "Always specify the wiring mode and transformer ratings prior to setting up setpoints and analog outputs."
  • "The maximum value for the product of the phase CT primary current and PT ratio is 57,500,000. If the product is greater, power readings are zeroed."

How to view Ct without changing it

The primary menu has three entries. In the manual's words, OPS is the "Main setup menu entry allowing to review setup options", and CHG is the "Main setup menu entry allowing to change setups". These steps follow the manual's menu instructions.

  1. From the data display, press and release SELECT to open the primary menu.
  2. Press SELECT to activate the OPS window, then press ENTER to enter the main menu.
  3. Press SELECT to highlight the upper item, use the UP and DOWN buttons until bASc appears, then press ENTER.
  4. Press SELECT to highlight the middle window, then use UP and DOWN until Ct appears. The lower item shows the present value.
  5. Press ESC to leave.

The manual's step to change a parameter starts with "Press the SELECT button to highlight the lower item". SATEC Australia's advice: when you are only reviewing, don't press SELECT at step 4 once Ct is showing. Don't enter the CHG menu.

SATEC Australia's advice: write down the Ct value together with the CT nameplate rating and the meter's serial number. Any change to Ct should be an authorised change made by the person responsible for the meter.

If Ct doesn't match the CT nameplate

SATEC Australia's advice: if the Ct value is not the primary rating shown on the CT nameplate, don't change it during a review. Contact SATEC Australia or the person responsible for the meter, and include these details:

  • the meter's serial number and rear label, including the current input
  • the CT nameplate rating, and how many CTs are fitted
  • the values of Ct, Pt, Pt.F, U.SEC and ConF, as viewed through OPS
  • the firmware versions shown in the display setup menu (diSP)
  • a few current readings from the display, with the time they were taken

Other current-input versions

Addenda to the PM17x manuals cover versions with external current sensors, and they give different rules for the CT primary current.

  • RS5 and HACS. The External CT addendum says these versions "require setup of "CT Primary current"", and that "This value must be set to In: Where: In = 1/2 Imax (RATING)". It adds: "If it is not direct connection this value should be multiplied to the Primary CT transformer ratio."
  • SC50. The add-on series addendum says: "In device general set-up CT primary current should be multiplied by 10 for PM172/174/175SC50 power meters, power quality analyzers, and fault recorders."

Both addenda say they apply to the PM175 manual BG0415 Rev. A6, the revision before the current one.

SATEC Australia's advice: if the rear label shows HACS, RS5, SC5 or SC50, don't use this page. Contact SATEC Australia with the serial number.

Not covered here

  • The TFT touch-screen display. The PM175 manual's menu instructions are for the LED display, and the manual doesn't mention a touch screen. SATEC Australia's advice: contact SATEC Australia for touch-screen documentation.
  • Changing the setting. This page covers viewing only.
  • PAS. For which PAS actions read from or write to the meter, see Which PAS actions read from, save locally or write to my meter?

Manual

PM175 Installation and Operation Manual BG0415 Rev. A7. The Basic setup table is on page 44, PM175 manual, page 44. The menu instructions are on pages 41 to 43.

Need help?

Contact SATEC Australia technical support through our contact form.

Source review: 22 November 2026.