EDL175 · BATTERY
Looking after the EDL175 battery

The EDL175 carries its own rechargeable battery, which keeps the logger running through a short supply interruption. This article covers what the manufacturer's manual says about charging, storage and the indicator lights, and who to contact if the battery no longer holds a charge.
What the battery does
The battery inside the case keeps the EDL175 running when its AC supply is disconnected for a short interval, such as a power failure during a survey. It is rated 12 V, 1.2 Ah. Section 3.1 of the manual says a full charge runs the instrument for at least 20 minutes; the specifications give 20 minutes.
The manual also says an external 12 V battery can be connected at the battery measurement sockets for longer use without supply, but it gives no connection details. SATEC Australia's advice: contact SATEC Australia before connecting an external battery.
Charging
- Connect the EDL175 to an AC supply to charge it. The manual recommends 8 hours to recharge a fully discharged battery.
- With AC connected and the slide switch OFF, the battery charges rapidly. When AC is connected and the display is on, it charges slowly.
- Do not let the battery discharge completely.
- The manual's own warning: turn the slide switch OFF when the EDL175 is not in use.
Reading the indicator lights
| Green AC LED | Red BAT LED | Red LOW BAT LED | PM175 display | What it means |
|---|---|---|---|---|
| On | Off | Off | On | Running from the AC supply; battery charging slowly |
| Off | On | Off | On | Running from the internal battery |
| Off | On | Blinking, with buzzer | On | Battery needs charging. The EDL175 will switch itself off within 5 minutes |
| On | Off | Off | Off | AC connected with the slide switch OFF; battery charging rapidly |
Storing the EDL175
The manual gives the longest time a stored battery can go before it needs recharging. The warmer the store, the shorter the time:
| Storage temperature | Time before recharging is required |
|---|---|
| +10 °C | more than 18 months |
| +20 °C | 14 months |
| +30 °C | 8 months |
| +40 °C | 5 months |
The meter inside: clock and recorded data
The EDL175 is built around a PM175 power meter. The PM175 manual describes a battery-backed internal clock and battery-backed recording memory. It also describes a backup battery status on the status display, shown as "nor" (normal) or "Lo" (low), and a diagnostic 13, "Low battery", for which it says "Battery replacement is required".
Neither the PM175 manual nor the EDL175 manual gives the type, service life or a replacement procedure for that backup battery.
If the battery no longer holds a charge
The EDL175 manual gives no procedure for replacing the supply battery or its fuse, and the PM175 manual gives none for the meter's backup battery. The EDL175 manual states that the instrument "may be opened only by a duly authorized representative of the manufacturer", and that the warranty is on a return-to-factory basis.
SATEC Australia's advice: do not open the instrument. Contact SATEC Australia with the EDL175's serial number through our contact form.
Flexible current sensors
The EDL175 works with FLEX and clamp current sensors that have a voltage output of up to 3 V RMS, connected through the optional FLEX connector cable. The EDL175 manual does not say whether these sensors need their own battery. SATEC Australia's advice: contact SATEC Australia to check the sensors supplied with your EDL175.
Source review: 22 November 2026.