# PM135 clock backup and the TOU module battery

> How the PM135 keeps time when its supply is off, which arrangements have a battery, and who to contact about the TOU module's CR1632 battery.

Source page: https://satec-global.com.au/support/articles/pm135-clock-backup-tou-module-battery/

---

PM135 · CLOCK BACKUP

# PM135 clock backup and the TOU module battery

*[PM135 and its TOU add-on module, from its manufacturer manual]*PM135 and TOU add-on module, manufacturer manual page 35. Identification illustration only.

Having a clock does not by itself mean a PM135 has a battery. In the PM135 manual (BG0518 Rev. A4), the regular version holds its clock for only seconds with the supply off, using a super-capacitor. The optional TOU module adds a battery-backed clock, with a CR1632 battery. The manual also mentions a V2 version of the meter with its own battery-backed clock. This article explains each arrangement, how to check the meter's clock, and who to contact about the TOU module battery.

## The regular PM135: a super-capacitor

The manual's specifications describe the regular PM135 clock as a "Super Cap backup clock": a super-capacitor, not a battery. The manual gives two figures for how long it keeps time with the supply off: 20 seconds in the feature list and a typical 30 seconds in the specifications.

The manual gives no replacement or maintenance procedure for the super-capacitor.

## The optional TOU module: a battery-backed clock

The TOU module is an optional expansion module, described as a battery-operated clock unit with four digital inputs. Its clock is battery-backed, with a typical retention of 36 months with the supply off (three years in the manual's options list). The cell is a CR1632, shown in the manual as a 3 V lithium battery.

The manual says the TOU battery-backed clock unit is "highly recommended" when the meter uses time-scheduled tariff rates. Otherwise, it says, a long power outage may cause the meter clock to lose time, so the tariff counters would not comply with the calendar schedule.

In the manual's drawings, the module's label has a box marked "MOD. TOU", and the module case is marked "TOU MODULE". The front-panel Device Info display has an I/O module type field, but the manual does not show what it reads with a TOU module fitted.

## The V2 version

The manual also describes a "PM135 version V2" whose standard meter clock is battery-backed, with a typical retention of 36 months (three years in the feature list). It does not give that battery's type, life or a replacement procedure. It does not say how to tell a regular PM135 from a V2: it names no label, order code, serial-number range or display field for this, and the current datasheets do not list a V2 option.

**SATEC Australia's advice:** contact SATEC Australia with the meter's serial number to find out which clock version it has. The Device Info display shows the serial number, model, and firmware and bootloader versions.

The retention figures above are how long the clock keeps time with the supply off. The manual does not give a service life for either battery.

## Checking the clock

- After a supply interruption, check the meter's date and time. The front display shows them at the left of its status bar, and the PAS RTC dialog shows the time in the meter.
- The manual describes setting the clock from the front panel's Real Time Clock menu or from PAS.
- The manual lists a device diagnostic, "RTC fault", meaning "The clock time has been lost".

## Replacing the TOU module battery

The manual includes a "TOU – Battery Replacement" procedure for the CR1632. It warns that only qualified personnel familiar with the instrument and its associated electrical equipment must carry it out. The procedure involves removing the TOU module from the meter and opening the module case.

The manual leaves some points open:

- The procedure does not state whether the meter must be powered off. Elsewhere, the manual says all incoming power sources must be turned off before any work on the instrument, and before installing an I/O or communication module. It also says: "Do not open the instrument under any circumstances when it is connected to a power source."
- The manual's warranty states that "only a duly authorised representative of the manufacturer may open your instrument", and that it should only be opened in a fully anti-static environment.
- The text and the figure describe the cell's orientation in different terms. The text says the "(+) battery pole is toward the battery holder"; the figure marks the positive side "On the top".
- There is no step for refitting the module, and the manual does not say whether the clock needs setting again, or whether stored data is affected.

**SATEC Australia's advice:** do not remove or open the TOU module while the meter is connected to any power source. Before replacing the TOU module battery, contact SATEC Australia with the meter's serial number through our [contact form](https://satec-global.com.au/contact/).

Source review: 22 November 2026.
